NI 5122/5124 Calibration Procedure
14
ni.com
Table 3 for channel 0, then repeat the procedures for channel 1. The
Calibration Test Limits and Published Specifications for vertical accuracy
are shown in Table 4 (NI 5122) and Table 5 (NI 5124).
To verify vertical gain accuracy, complete the procedures described in the
section for each of the 39 iterations listed in
Table 3 for channel 0, then repeat the procedures for channel 1. The
Calibration Test Limits and Published Specifications for vertical gain are
shown in Table 6 (NI 5122) and Table 7 (NI 5124).
Table 3.
NI 5122/5124 Input Parameters for Vertical Offset Accuracy
and Vertical Gain Accuracy Verification
Iteration
Input
Impedance
Max Input Frequency (Hz)
Range
NI 5122
NI 5124
1
50
Ω
100,000,000
150,000,000
0.2
2
50
Ω
100,000,000
150,000,000
0.4
3
50
Ω
100,000,000
150,000,000
1
4
50
Ω
100,000,000
150,000,000
2
5
50
Ω
100,000,000
150,000,000
4
6
50
Ω
100,000,000
150,000,000
10
7
50
Ω
35,000,000
60,000,000
0.2
8
50
Ω
35,000,000
60,000,000
0.4
9
50
Ω
35,000,000
60,000,000
1
10
50
Ω
35,000,000
60,000,000
2
11
50
Ω
35,000,000
60,000,000
4
12
50
Ω
35,000,000
60,000,000
10
13
50
Ω
20,000,000
20,000,000
0.2
14
50
Ω
20,000,000
20,000,000
0.4
15
50
Ω
20,000,000
20,000,000
1
16
50
Ω
20,000,000
20,000,000
2
17
50
Ω
20,000,000
20,000,000
4
18
50
Ω
20,000,000
20,000,000
10
19
1 M
Ω
100,000,000
150,000,000
0.2
20
1 M
Ω
100,000,000
150,000,000
0.4