background image

Caution

 

To prevent data loss and to maintain the integrity of your Ethernet

installation, do not use a cable longer than 100 m.

If you need to build your own cable, refer to the 

Cabling

 section for more information about

Ethernet cable wiring connections.

Note

 

You will configure the IP settings as described in the 

Configuring IP Settings

section, after connecting power and powering the chassis on.

The host computer communicates with the chassis over a standard Ethernet connection. If the
host computer is on a network, you must configure the chassis on the same subnet as the host
computer. If neither the host computer nor the chassis is connected to a network, you can
connect the two directly using a crossover cable.

If you want to use the chassis on a subnet other than the one the host computer is on, first
connect the chassis on the same subnet as the host computer. Use DHCP to assign an IP
address or reassign a static IP address for the subnet where you want it to be and physically
move it to the other subnet.

Wiring Power to the Chassis

The NI 9148 requires an external power supply that meets the specifications in the 

Power

Requirements

 section.

The NI 9148 filters and regulates the supplied power and provides power for all of the I/O
modules. The NI 9148 has one layer of reverse-voltage protection. Complete the following
steps to connect a power supply to the chassis.
1.

Ensure that the power supply is turned off.

2.

Connect the positive lead of the power supply to the V terminal of the COMBICON
power connector shipped with the NI 9148, and tighten the terminal screw. The following
figure shows the terminal screws, which secure the wires in the screw terminals, and the
connector screws, which secure the power connector on the controller.

Figure 11.

 COMBICON Power Connector

C

NC

C

V

2

1

2

1. Terminal Screw
2. Connector Screws

3.

Connect the negative lead of the power supply to the C terminal of the power connector
and tighten the terminal screw.

10

  | 

ni.com

  | 

NI 9148 User Manual

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for 9148

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...the NI 9148 Figure 1 NI 9148 Front Panel NI 9148 POWER USER FPGA1 STATUS SAFE MODE IP RESET INPUT 19 30 V 20 W MAX LINK 10 100 V C NC C DO NOT SEPARATE CABLES WHEN ENERGIZED IN HAZARDOUS LOCATIONS 2...

Page 3: ...hed off or the area is known to be nonhazardous Caution Substitution of components may impair suitability for Class I Division 2 Caution For Division 2 and Zone 2 applications install the system in an...

Page 4: ...e Approved for marine shipboard applications To verify Lloyd s Register certification for a product visit ni com certification and search for the LR certificate or look for the Lloyd s Register mark o...

Page 5: ...n any orientation on a 35 mm DIN rail or on a panel Use the DIN rail mounting method if you already have a DIN rail configuration or if you need to be able to quickly remove the chassis Use the panel...

Page 6: ...he serial number from the back of the chassis You will be unable to read the serial number after you have mounted the chassis Caution Make sure that no I O modules are in the chassis before mounting i...

Page 7: ...PUT 19 30 V 20 W MAX LINK 10 100 V C NC C 2 Fasten the NI 9905 panel to the wall using the screwdriver and screws that are appropriate for the wall surface The maximum screw size is M4 or number 8 Cau...

Page 8: ...ing of the DIN rail clip Figure 8 One Edge of the DIN Rail Inserted in a Clip 3 2 1 1 DIN Rail Clip 2 DIN Rail 3 DIN Rail Spring 3 Press down firmly on the chassis to compress the spring until the cli...

Page 9: ...n 1 Make sure that no I O side power is connected to the I O module If the system is in a nonhazardous location the chassis power can be on when you install I O modules 2 Align the I O module with an...

Page 10: ...a C Series I O module from the chassis 1 Make sure that no I O side power is connected to the I O module 2 Squeeze the latches on both sides of the module and pull the module out of the chassis Conne...

Page 11: ...the subnet where you want it to be and physically move it to the other subnet Wiring Power to the Chassis The NI 9148 requires an external power supply that meets the specifications in the Power Requi...

Page 12: ...m flash memory to the FPGA when the chassis powers on Autoload VI on device reboot Loads the FPGA bit stream from flash to the FPGA when you reboot the chassis either with or without cycling power Con...

Page 13: ...e Notes Configuring DIP Switches Both DIP switches are in the OFF position when the chassis is shipped from National Instruments Figure 12 DIP Switches 1 2 OFF SAFE MODE IP RESET SAFE MODE Switch Safe...

Page 14: ...define the USER FPGA1 LED to meet the needs of your application The following table lists the USER FPGA1 LED indicators Table 3 User LEDs LED LED Color Description USER FPGA1 Green Yellow Use the LabV...

Page 15: ...ce may be configured for DHCP but unable to get an IP address because of a problem with the DHCP server Check the network connection and try again If the problem persists contact National Instruments...

Page 16: ...ion built into the product if the product is damaged in any way If the product is damaged return it to NI for repair Network Network interface 10BaseT and 100BaseTX Ethernet Compatibility IEEE 802 3 C...

Page 17: ...ry is for measurements of voltages from specially protected secondary circuits Such voltage measurements include signal levels special equipment limited energy parts of equipment circuits powered by r...

Page 18: ...cations Note Group 1 equipment per CISPR 11 is any industrial scientific or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection an...

Page 19: ...ndensing Storage humidity IEC 60068 2 56 5 RH to 95 RH noncondensing Pollution Degree IEC 60664 2 Maximum altitude 2 000 m Indoor use only 1 Shock and Vibration To meet these specifications you must p...

Page 20: ...I is committed to designing and manufacturing products in an environmentally responsible manner NI recognizes that eliminating certain hazardous substances from our products is beneficial to the envir...

Page 21: ...ngineers Visit ni com services for NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your NI product Product registration facilitates tech...

Page 22: ...NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export da...

Page 23: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Reviews: