4.
Select settings for the chassis in the
IP Settings
section, then click
Apply
.
Note
For information about configuring network settings, refer to the
Configuring Network Settings book of the
MAX Remote Systems Help
. In
MAX, click
Help
»
Help Topics
»
Remote Systems
. On the
Contents
tab,
browse to
LabVIEW Real-Time Target Configuration
»
Configuring
Network Settings
.
5.
When you click
Apply
, you are prompted to reboot the chassis for the changes to take
effect. Click
Yes
. You can also reboot the chassis by right-clicking the name under
Remote Systems
and selecting
Reboot
.
6.
After rebooting, the chassis appears under
Remote Systems
with the assigned name.
Expand the chassis and select
Software
.
7.
Click
Add/Remove Software
in the toolbar to launch the LabVIEW Real-Time Software
Wizard.
8.
Install the Recommended Software Set that appears in the LabVIEW Real-Time Software
Wizard.
For more information about configuring the chassis in MAX, refer to the
MAX Help
. For
information about installing and using LabVIEW FPGA, refer to the
LabVIEW FPGA Module
Release and Upgrade Notes
.
Configuring DIP Switches
Both DIP switches are in the OFF position when the chassis is shipped from National
Instruments.
Figure 12.
DIP Switches
1
2
OFF
SAFE MODE
IP RESET
SAFE MODE Switch
Safe mode is for troubleshooting, updating configuration, and installing software. Keep the
SAFE MODE switch in the OFF position during normal operation. If the switch is in the ON
position at startup, the NI 9148 launches only the essential services required for updating
configuration and installing software.
Push the SAFE MODE switch to the ON position if the software on the chassis is corrupted.
Even if the switch is not in the ON position, if there is no software installed on the chassis, the
chassis automatically boots into safe mode. The SAFE MODE switch must be in the ON
position to reformat the drive on the chassis. Refer to the
Measurement & Automation
Explorer Help
for more about installing software and reformatting the drive.
IP RESET Switch
Push the IP RESET switch to the ON position and reboot the chassis to reset the IP address
and other TCP/IP settings of the chassis to the factory defaults. Refer to the
Troubleshooting
12
|
ni.com
|
NI 9148 User Manual