CALIBRATION PROCEDURE
NI 9218
2-Channel, 24-Bit, Universal Analog Input Module
This document contains information for calibrating the National Instruments 9218. For more
information on calibration, visit
ni.com/calibration
.
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 4
Initial Setup............................................................................................................................... 5
Verification............................................................................................................................... 5
Current Accuracy Verification, 20 mA mode .................................................................. 6
Voltage Accuracy Verification, 60 V mode ..................................................................... 7
Voltage Accuracy Verification, 16 V mode ..................................................................... 9
Voltage Accuracy Verification, 65 mV mode .................................................................. 10
Voltage Accuracy Verification, IEPE 5 V mode.............................................................. 12
Bridge Accuracy Verification, 22 mV/V mode................................................................ 14
Voltage Accuracy Adjustment, 16 V mode...................................................................... 17
Voltage Accuracy Adjustment, 60 V mode...................................................................... 18
Voltage Accuracy Adjustment, 65 mV mode................................................................... 19
Current Accuracy Adjustment, 20 mA mode ................................................................... 20
Voltage Accuracy Adjustment, IEPE 5 V mode .............................................................. 20
Bridge Accuracy Adjustment, 22 mV/V mode ................................................................ 21
EEPROM Update ..................................................................................................................... 23
Reverification ........................................................................................................................... 23
Accuracy Under Calibration Conditions .................................................................................. 23
Worldwide Support and Services ............................................................................................. 24
Pinout Signals ........................................................................................................................... 25
Software
Calibrating the NI 9218 requires the installation of NI-DAQmx 15.1or later on the calibration
system. You can download NI-DAQmx from
ni.com/downloads
. NI-DAQmx supports
LabVIEW, LabWindows
™
/CVI
™
, C/C++, C#, and Visual Basic .NET. When you install
NI-DAQmx you only need to install support for the application software that you intend to use.