12
|
ni.com
|
NI 9218 Calibration Procedure
Voltage Accuracy Verification, IEPE 5 V mode
Complete the following procedure to determine the As-Found status of the NI 9218.
1.
Connect the NI 9218 to the calibrator as shown in Figure 5.
Figure 5.
Voltage Accuracy Verification Connections
2.
Configure calibrator according to Table 14.
Note
The test limits listed in Table 14 are derived using the values in Table 32.
Note
The uncertainty of the calibration standard is not included in these limits.
3.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
4.
Acquire and calculate the AC magnitude.
a.
Create an AI voltage task on the NI 9218 according to Table 15.
Table 14.
NI 9218 Verification Test Limits for Accuracy
Test Point
2-Year Limits
Signal Type
Frequency
(kHz)
Amplitude
(Vrms)
Lower Limit
(Vrms)
Upper Limit
(Vrms)
Sine wave
1
1
0.998
1.002
3
2.994
3.006
Table 15.
NI 9218 Configuration for Phase Matching Verification
Min (V)
Max (V)
Scaled Units
Channel Configuration
-5
5
Volts
AC Coupled
AI+
AI–
NI 99
8
2
C
a
li
b
r
a
tor
–
+
AI+
AI–
NI 99
8
2
NI 921
8
D
S
UB/LEMO
CH 0
D
S
UB/LEMO
CH 1