NI 9218 Calibration Procedure
Test Conditions
The following setup and environmental conditions are required to ensure the NI 9218
meets
published specifications.
•
Keep connections to the device as short as possible. Long cables and wires act as antennas,
picking up extra noise that can affect measurements.
•
Verify that all connections to the device are secure.
•
Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to
eliminate noise and thermal offsets.
•
Maintain an ambient temperature of 23 ±5 °C. The device temperature will be greater than
the ambient temperature.
•
Keep relative humidity below 80%.
•
Allow a warm-up time of at least 10 minutes to ensure that the NI 9218 measurement
circuitry is at a stable operating temperature.
•
Only one of each NI 9982, NI 9983, and NI 9987 is sufficient for this calibration procedure.
Discrete
Resistors
—
Bridge,
±22 mV
mode
Bridge,
±22 mV
mode
TCR:
<25
ppm/°C
Resistor values:
10 k
Ω
(2x), 249
Ω
cDAQ
Chassis
NI cDAQ-9178
All
All
—
Measurement
Adapter
NI 9982 (x2)
±16 V,
±65 mV,
Bridge
±22 mV/V,
IEPE ±5 V
mode
±60 V, ±16 V,
±65 mV,
Bridge
±22 mV/V,
IEPE ±5 V
mode
—
NI 9983 (x2)
±20 mA
mode
—
—
NI 9987 (x2)
±60 V mode
—
—
Table 1.
Recommended Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
Verification
Adjustment