16
|
ni.com
|
NI 9218 Calibration Procedure
7.
Acquire and average samples.
a.
Create an AI bridge with excitation task on the NI 9218 according to Table 18.
b.
Configure the AI Bridge with excitation task timing according to Table 19.
c.
Start the task.
d.
Record the DMM1 measurements as V
DMM1
.
e.
Record the DMM2 measurement as V
DMM2
.
Note
Perform the DMM measurements and the NI 9218 measurements as
simultaneously as possible in order to reduce error from drift in the resistors.
f.
Average the NI 9218 readings and record the average as Ratio
channel
.
g.
Perform the following calculation using the two DMM measurements.
h.
Perform the following calculation using the recorded Ratio
ref
and Ratio
channel
values.
Accuracy
=
Ratio
channel
-
Ratio
ref
i.
Compare the result of step h to the limits in Table 17.
j.
Stop and clear the task.
8.
Repeat steps 2 through 7 for each test point in Table 17.
9.
Repeat steps 1 through 8 for each excitation value specified in Table 17.
10. Disconnect the DMMs and resistors from the NI 9218.
11. Repeat steps 1 through 10 for each channel on the NI 9218.
Table 18.
NI 9218 Configuration for Accuracy Verification
Min
(V/V)
Max
(V/V)
Channel
Configuration
Excitation
Value (V)
Excitation
Source
Bridge
Configuration
-0.022
0.022
DC Coupled
2
Internal
Full-Bridge
3.3
Table 19.
NI 9218 Timing Configuration
Sample Mode
Samples Per Channel
Rate (kHz)
Finite Samples
512 k
51.2
Ratio
ref
V
DMM
2
V
DMM
1
------------------
=