background image

 

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. 
Refer to the 

Terms of Use

 section on 

ni.com/legal

 for more information about National 

Instruments trademarks. Other product and company names mentioned herein are trademarks or trade 
names of their respective companies. For patents covering National Instruments products, refer to the 
appropriate location: 

Help»Patents

 in your software, the 

patents.txt

 file on your CD, or 

ni.com/patents

.

© 2007 National Instruments Corporation. All rights reserved.

372251A-01

Jun07

Where to Go for Support

The National Instruments Web site is your complete resource for technical 
support. At 

ni.com/support

 you have access to everything from 

troubleshooting and application development self-help resources to email 
and phone assistance from NI Application Engineers.

National Instruments corporate headquarters is located at 
11500 North Mopac Expressway, Austin, Texas, 78759-3504. 
National Instruments also has offices located around the world to help 
address your support needs. For telephone support in the United States, 
create your service request at 

ni.com/support

 and follow the calling 

instructions or dial 512 795 8248. For telephone support outside the United 
States, contact your local branch office: 

Australia 1800 300 800, Austria 43 662 457990-0, 
Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599, 
Canada 800 433 3488, China 86 21 5050 9800, 
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00, 
Finland 385 (0) 9 725 72511, France 01 57 66 24 24, 
Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737, 
Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400, 
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, 
Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466, 
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, 
Poland 48 22 3390150, Portugal 351 210 311 210, Russia 7 495 783 6851, 
Singapore 1800 226 5886, Slovenia 386 3 425 42 00, 
South Africa 27 0 11 805 8197, Spain 34 91 640 0085, 
Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151, 
Taiwan 886 02 2377 2222, Thailand 662 278 6777, 
Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545

Summary of Contents for 9237

Page 1: ...ler specific instructions and detailed function explanations You can install these help files when you install NI DAQmx on the calibration computer Contents Conventions 2 Software 2 Documentation 3 Calibration Interval 3 Test Equipment 4 Test Conditions 4 Calibration Procedure 5 Calibration Process Overview 5 Initial Setup 5 Verification Procedure 6 Specifications 12 Test Limits 12 Using the Table...

Page 2: ...ng examples and syntax examples This font is also used for the proper names of disk drives paths directories programs subprograms subroutines device names functions operations variables filenames and extensions monospace italic Italic text in this font denotes text that is a placeholder for a word or value that you must supply Software Install NI DAQmx 8 1 or later on the calibration computer NI D...

Page 3: ...software and NI DAQmx supported DAQ devices and how to confirm that your device is operating properly To access this guide select Start All Programs National Instruments NI DAQ DAQ Getting Started Guide Note The documents above are installed with NI DAQmx You can also download the latest versions from the NI Web site at ni com manuals NI 9237 Operating Instructions This document describes how to u...

Page 4: ...nt temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the measurement circuitry is at a stable operating temperature Table 1 Recommended Equipment Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable use a high precision voltage source with an accuracy of at least 10 ppm and an output impedance of less...

Page 5: ...justment The first two steps are explained in the following sections Initial Setup You must configure the device in Measurement Automation Explorer MAX to communicate with NI DAQmx Complete the following steps to configure a device in MAX 1 Install the NI DAQmx driver software 2 Make sure that no power is connected to the module terminals If the system is in a nonhazardous location the chassis pow...

Page 6: ...lowing steps to test the performance of the device 1 Make the following connections on the NI 9949 terminal block Insert one leg of each 350 Ω resistor into AI Insert the second leg of one 350 Ω resistor into EX Insert the second leg of the other 350 Ω resistor into EX Connect RS to EX Connect RS to EX Connect the positive input of the DMM to RS Connect the negative input of the DMM to RS Connect ...

Page 7: ...cription 1 SC Shunt calibration 2 AI Positive analog input 3 AI Negative analog input 4 RS Positive remote sense input 5 RS Negative remote sense input 6 EX Positive excitation output 7 EX Negative excitation output 8 T TEDS DATA 9 T TEDS Return 10 SC Shunt calibration These signals are shared by all four RJ 50 connectors on the NI 9237 EX RS RS EX DMM AI Calibrator Guard AI ...

Page 8: ...ing only the values used in your application 3 If you use C function calls create a task using DAQmxCreateTask as shown in the following table If you use LabVIEW skip this step The task is created in step 4 in LabVIEW LabVIEW Block Diagram NI DAQmx Function Call LabVIEW does not require this step Call DAQmxCreateTask with the following parameters taskName AIVerificationTask taskHandle taskHandle ...

Page 9: ...agram NI DAQmx Function Call Call DAQmxCreateAIVoltageChan WithExcit with the following parameters taskHandle taskHandle physicalChannel dev1 ai0 nameToAssignToChannel myVoltageChannel terminalConfig DAQmx_Val_Cfg_Default minVal 0 025 maxVal 0 025 units DAQmx_Val_Volts bridgeConfig DAQmx_Val_FullBridge voltageExcitSource DAQmx_Val_Internal voltageExcitVal 2 5 3 3 5 0 or 10 0 useExcitforScaling TRU...

Page 10: ... DAQmx Start Task VI as shown in the following table LabVIEW Block Diagram NI DAQmx Function Call CallDAQmxCfgSampClkTiming with the following parameters taskHandle taskHandle source NULL rate 50000 activeEdge DAQmx_Val_Rising sampleMode DAQmx_Val_FiniteSamps sampsPerChan 10000 LabVIEW Block Diagram NI DAQmx Function Call Call DAQmxStartTask with the following parameter taskHandle taskHandle ...

Page 11: ...p 7 to the upper limit and lower limit values you calculated in step 9 If the result is between these values the device passes the test 11 Clear the acquisition using the DAQmx Clear Task VI as shown in the following table 12 Repeat steps 2 through 11 for all voltage values all excitation values and all channels 13 Disconnect the calibrator and the DMM from the device LabVIEW Block Diagram NI DAQm...

Page 12: ... Setting Value is the voltage value to enter in the DAQmx Create Virtual Channel VI in step 4 Calibrator Output The Calibrator Output is the voltage value that you source into the NI 9237 for verification purposes This value is broken down into two columns Location and Value Location refers to where the value fits within the test range Value refers to the voltage value to be verified Max refers to...

Page 13: ... Output Range V V Test Point Value V V 1 Year Absolute Error V V Location Value V Minimum Maximum 2 5 Max 0 055 0 025 0 025 0 022000 0 000063 2 5 Mid 0 000 0 025 0 025 0 000000 0 000033 2 5 Min 0 055 0 025 0 025 0 022000 0 000063 3 3 Max 0 073 0 025 0 025 0 022000 0 000036 3 3 Mid 0 000 0 025 0 025 0 000000 0 000010 3 3 Min 0 073 0 025 0 025 0 022000 0 000036 5 0 Max 0 110 0 025 0 025 0 022000 0 0...

Page 14: ...onal Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your service request at ni com support and follow the calling instructions or dial 512 795 8248 For telephone support outside the United States contact your local branch office Australia 1800 300 800 Austria 43 662 457990 0 Belgium 32 0 2 757 0020 Brazil 5...