Test Equipment

The following table lists the equipment recommended for the performance verification
procedure. If the recommended equipment is not available, select a substitute using the
requirements listed.

Table 1.

 Recommended Equipment

Equipment

Recommended Model

Requirements

 

 

 

Frequency Standard NI PXIe-6674T,

NI PXIe-6614, or
Keysight 33250A

A highly stable clock source that is
able to generate 1 MHz, 5 V TTL
output with ≤±5 ppm total accuracy

Chassis

NI cDAQ-9178

Connection
Accessory

NI 9923

Test Conditions

The following setup and environmental conditions are required to ensure the NI 9361 meets
published specifications.