Test Equipment
The following table lists the equipment recommended for the performance verification
procedure. If the recommended equipment is not available, select a substitute using the
requirements listed.
Table 1.
Recommended Equipment
Equipment
Recommended Model
Requirements
Frequency Standard NI PXIe-6674T,
NI PXIe-6614, or
Keysight 33250A
A highly stable clock source that is
able to generate 1 MHz, 5 V TTL
output with ≤±5 ppm total accuracy
Chassis
NI cDAQ-9178
—
Connection
Accessory
NI 9923
—
Test Conditions
The following setup and environmental conditions are required to ensure the NI 9361 meets
published specifications.
2
|
ni.com
|
NI 9361 Calibration Procedure