3-2
|
ni.com
Chapter 3
Safety Information
-
Refer to
for information about internal cables.
Removing Power
Caution
To completely interrupt power to a single phase system, you
must
disconnect the AC power cable. Do
not
position equipment so that it is difficult
to disconnect the cable.
To completely interrupt power to a three phase system, you
must
switch off the Main
Breaker on the Power Entry Panel. Do
not
position equipment so that it is difficult to
access the main breaker.
Caution
You
must
press the Emergency Power Off (EPO) button, or turn the Main
Power Switch on the EPO to the off position, to fully disable the UPS output power
if the system includes the optional UPS.
Table 3-2.
Mid Power Configuration Power Cables
Part Number
Description
785718-01
AC, Single Phase IEC 60309, 2P+E, Receptacle,
to IEC 60309, 2P+E, Plug, 2.5 m
785719-01
AC, Single Phase IEC 60309, 2P+E, Receptacle,
to NEMA L6-30, 2.5 m
785735-01
IEC 60309, 2P+E, pin and sleeve (connector only)
Table 3-3.
High Power Configuration Power Cable
Part Number
Description
785717-01
AC, Three Phase IEC 60309, 3P+N+E, Receptacle,
to IEC 60309, 3P+N+E, Plug, 2.5 m
785734-01
IEC 60309, 3P+N+E, pin and sleeve (connector only)