Figure 22.
Bottom View with Dimensions in inches [millimeters]
2X 6–32
6.35 mm (0.250 in.)
11.43 mm
(0.450 in.)
11.43 mm
(0.450 in.)
65.19 mm
(2.566 in.)
65.19 mm
(2.566 in.)
38.10 mm
(1.50 in.)
Securing the NI CVS-1459 to a Mount
1.
Align the screw holes of the mounting bracket with the two holes on the back of the NI
CVS-1459.
2.
Insert two 6-32 screws and tighten them until they are secure (3.5 in-lb). Make sure the
heads of the screws are flush with the mounting bracket.
Clearance Requirements
The NI CVS-1459 installation must meet the following space and cabling clearance
requirements for optimum cooling:
•
Allow 76.2 mm (3.0 in.) on the top and bottom of the NI CVS-1459 for air circulation.
•
Allow 50.8 mm (2.0 in.) on the sides of the NI CVS-1459 for air circulation.
•
Allow enough space in front of the NI CVS-1459 to connect cables.
NI CVS-1459 User Manual
|