Chapter 2
Analog Input Timing/Control
DAQ-STC Technical Reference Manual
2-98
©
National Instruments Corporation
Figure 2-27.
START1 Delays, Synchronous Mode, Internal CONVERT
Figure 2-28.
START2 Delays, Synchronous Mode, Internal CONVERT
In the external CONVERT mode, the active edge of CONVERT_SRC that recognizes the
external trigger generates the output. Figure 2-29 shows the propagation delays for START1.
Figure 2-30 shows the propagation delays for START2.
Figure 2-29.
START1 Delays, Synchronous Mode, External CONVERT
START1
SI2 Source
PFI0/AI_START1
RTSI_TRIGGER<0..6>
RTSI_BRD<0..3>
Tpfi
Trtsi
Tbrd
START2
SI2 Source
PFI1/AI_START2
RTSI_TRIGGER<0..6>
RTSI_BRD<0..3>
Tpfi
Trtsi
Tbrd
START1
CONVERT_SRC
PFI0/AI_START1
RTSI_TRIGGER<0..6>
RTSI_BRD<0..3>
Tpfi
Trtsi
Tbrd