© National Instruments
|
4-5
Table 4-2 shows how signals are routed to the NI-PGIA on MIO X Series devices.
For differential measurements, AI 0 and AI 8 are the positive and negative inputs of differential
analog input channel 0. For a complete list of signal pairs that form differential input channels,
refer to the
I/O Connector Signal Descriptions
.
Table 4-2.
Signals Routed to the NI-PGIA on MIO X Series Devices
AI Ground-Reference
Settings
Signals Routed to the
Positive Input of the
NI-PGIA (Vin+)
Signals Routed to the
Negative Input of the
NI-PGIA (Vin-)
RSE
AI <0..207>
AI GND
NRSE
AI <0..15>
AI SENSE
AI <16..79>
AI SENSE 2
AI <80..143>
AI SENSE 3
AI <144..207>
AI SENSE 4
DIFF
AI <0..7>
AI <8..15>
AI <16..23>
AI <24..31>
AI <32..39>
AI <40..47>
AI <48..55>
AI <56..63>
AI <64..71>
AI <72..79>
AI <80..87>
AI <88..95>
AI <96..103>
AI <104..111>
AI <112..119>
AI <120..127>
AI <128..135>
AI <136..143>
AI <144..151>
AI <152..159>
AI <160..167>
AI <168..175>
AI <176..183>
AI <184..191>
AI <192..199>
AI <200..207>