Calibration Circuitry
The X Series analog inp
u
ts and o
u
tp
u
ts have calibration circ
u
itry to correct
gain and offset errors. Yo
u
can calibrate the device to minimize AI and AO
errors ca
u
sed by time and temperat
u
re drift at r
u
n time. No external
circ
u
itry is necessary; an internal reference ens
u
res high acc
u
racy and
stability over time and temperat
u
re changes.
Factory-calibration constants are permanently stored in an onboard
EEPROM and cannot be modified. When yo
u
self-calibrate the device, as
described in the
Device Self-Calibration
section of Chapter 1,
Getting
Started
, software stores new constants in a
u
ser-modifiable section of the
EEPROM. To ret
u
rn a device to its initial factory calibration settings,
software can copy the factory-calibration constants to the
u
ser-modifiable
section of the EEPROM. Refer to the
NI-DAQmx Help
or the
LabVIEW
Help
for more information abo
u
t
u
sing calibration constants.
For a detailed calibration proced
u
re for X Series devices, refer to the
B/E/M/S/X Series Calibration Procedure
by clicking
Manual Calibration
Procedures
on
ni.com/calibration
.
Cables and Accessories
Refer to the
PCI Express, PXI Express, and USB Mass Termination Device
Cables and Accessories
section for X Series PCI Express, PXI Express,
and USB Mass Termination device cable and accessory information.
Refer to the
USB Screw Terminal Device Accessories
section for X Series
USB Screw Terminal device accessory information.
PCI Express, PXI Express, and USB Mass Termination Device Cables
and Accessories
Caution
For compliance with Electromagnetic Compatibility (EMC) req
u
irements, this
prod
u
ct m
u
st be operated with shielded cables and accessories. If
u
nshielded cables or
accessories are
u
sed, the EMC specifications are no longer g
u
aranteed
u
nless all
u
nshielded cables and/or accessories are installed in a shielded enclos
u
re with properly
designed and shielded inp
u
t/o
u
tp
u
t ports.