Setting up the ECUTS
ECUTS-16000 Installing the Monitor and Monitor Arm
The ECUTS-16000 includes a monitor as well as a monitor arm with a keyboard tray. Use the following steps to install the monitor and
monitor arm.
1.
Refer to the Ergomart
TRS7000 Monitor Arm Installation Instructions
to attach the monitor arm to the ECUTS-16000 mounting cylinder.
2.
Connect the computer monitor to the ECUTS-16000 DisplayPort, USB, and AC power cables located on top of the test rack.
3.
Connect a USB keyboard and mouse to the monitor USB ports.
ECUTS-16001 Installing the Monitor and Monitor Tilter
The ECUTS-16001 includes a monitor as well as a monitor tilter. Use the following steps to install the monitor and monitor tilter.
1.
Refer to steps 2-5 in the Ergomart
WM9110S Monitor Wall Mount Installation Instructions
to attach the monitor to the ECUTS-16001 test
rack.
2.
Connect the computer monitor to the ECUTS-16001 DisplayPort, USB, and AC power cables located on the front of the test rack.
3.
Connect a USB keyboard and mouse to the monitor USB ports.
Installing the 9025TR Platform
If you have an ECUTS-16000 or ECUTS-16001 with a 9025TR mass interconnect, you must attach the accompanying 9025TR platform. Refer
to the Virginia Panel Corporation
9025 User Manual
to attach the 9025TR platform.
(Optional) Connecting to Facility Compressed Air
Use the following steps to connect the ECUTS to the facility compressed air.
The ECUTS does not ship ready for use with facility compressed air. Do not connect the ECUTS to the facility compressed air unless the
pneumatics have been configured.
Equipment
•
6 mm outer diameter air hose
1.
Connect a 6 mm outer diameter air hose to the pneumatic air port.
2.
Turn on the facility compressed air connected to the ECUTS.
Related Information
Bringing Up the ECUTS
Powering On the ECUTS
Notice
To avoid damage to the system, verify that an interchangeable test adapter (ITA) or field diagnostic tester (FDT) is not
connected to the ECUTS before powering on the system.
Equipment
•
Handheld DMM
•
LAN/Ethernet cable
1.
Verify that the main circuit breaker is in the off position on the power entry panel.
2.
(Optional for ECUTS-16001) Connect the ECUTS ground lug, located near the power receptacle on the rear of the ECUTS, to the facility
earth ground.
Note
The ECUTS ground lug connection requires 2.1 mm
2
(14AWG) copper wire.
3.
(Optional for ECUTS-16001) Verify that there is electrical continuity between the ECUTS ground lug and the facility earth ground using
the handheld DMM.
4.
Connect a network cable to the Ethernet port on the rear of the ECUTS.
5.
Confirm that the primary emergency power off (EPO) button is not engaged.
6.
Open the rear cabinet door.
7.
Verify that all PDU circuit breakers, including the main breaker and CB1 to CB3, are in the On position.
8.
Verify that the CONTROL switch is set to REMOTE.
9.
Close and lock the rear cabinet door.
10. Push in the brake lock on the ECUTS caster wheels to secure the test rack.
11. Connect the main AC power cable plug to the ECUTS power inlet.