32. Install the USB-6363, if present.
a)
Screw in the screws securing the bottom retention bracket to the right-side TRC panel using the #2 Phillips screwdriver.
b)
Insert the USB-6363 into the bottom retention bracket.
c)
Attach the top retention bracket by screwing in the two screws using the #2 Phillips screwdriver.
d)
Connect the cables to the USB-6363.
33. Screw in the screws securing the left-side PXIe-4112 auxiliary power supply bracket to the test rack using the #2 Phillips screwdriver.
34. Install the PXIe-4112 auxiliary power supplies on the bracket.
a)
Connect the AC power cable to the auxiliary power supply.
b)
Connect the auxiliary power input connector to the PXIe-4112.
c)
Connect the output connectors to the PXIe-4112.
35. Replace the filler panels on the front of the test rack.
36. Sit the left and right panels back into the test rack.
37. Turn the four quarter-turn locks on the left and right panels of the test rack using the flathead screwdriver.
38. Refer to
and complete the steps listed.
Configuring the SLSC Chassis
1.
Reset the SLSC chassis to the default factory settings.
a)
Press the Reset button on the rear of the SLSC for five seconds.
b)
Release the Reset button
c)
Press the Reset button for an additional five seconds.
2.
Sign in to the ECUTS.
3.
Launch NI MAX on the PXI controller and navigate to
My System
»
Devices and Interfaces
»
Network Devices
.
4.
Select the SLSC chassis from the Network Devices menu.
5.
Navigate to
Network Settings
.
6.
Click
Log In
.
7.
Login using the following credentials:
•
Username:
admin
•
Password: leave blank
8.
Select
Static
from the
Configure IPv4
drop-down menu.
9.
Enter
169.254.160.10
in the IPv4 Address text box.
10. Enter
255.255.0.0
in the Subnet Mask text box.
11. Click
Save
.
12. Select
Yes
when asked to reboot the SLSC chassis.
Replacing the PDU
(ECUTS-16000) Replacing the PDU
Equipment
•
#1 Phillips screwdriver
•
#2 Phillips screwdriver
•
Flathead screwdriver
1.
Power off the ECUTS-16000.
2.
Remove the four screws from each filler panel above the mass interconnect using the #2 Phillips screwdriver.
3.
Remove the filler panels.
4.
Open the rear cabinet door.
5.
Pull the spring pin attached to the inside hinge of the rear cabinet door down until it detaches from the test rack.
6.
Tilt and remove the rear cabinet door from the test rack.
7.
Turn the two quarter-turn locks on the left and right panels of the test rack using the flathead screwdriver.
8.
Remove the left and right panels of the test rack.
9.
Remove the USB-6363 from the left-side TRC bracket, if present.
a)
Disconnect the VHDCI cables from the USB-6363.
b)
Disconnect the power cable from the USB-6363.
c)
Remove the two screws on the USB-6363 top retention bracket using the #2 Phillips screwdriver.
d)
Remove the USB-6363 from the test rack.
ECUTS-16000/16001 User Manual
|
© National Instruments Corporation
|
69