FP-AI-100 and cFP-AI-100
12
ni.com
Current Inputs
The following input signal ranges are software selectable by
channel.
Input impedance................................ 100
Ω
Overcurrent protection
......................±
30 mA
Effective resolution ........................... 11.5 bits
Signal input bandwidth ..................... 160 Hz
Physical Characteristics
Indicators .......................................... Green
POWER
and
READY
indicators;
one red
Out of Range
indicator
Weight
FP-AI-100................................... 145 g (5.1 oz)
cFP-AI-100................................. 115 g (4.0 oz)
Power Requirement
Power from network module ............ 400 mW
Current
Input Range
Typical
Offset Error at
15 to 35
°
C
Maximum
Offset Error at
–40 to 70
°
C
0–20 mA
4–20 mA
±
20 mA
12
µΑ
12
µΑ
23
µΑ
20
µΑ
20
µΑ
40
µΑ
Current
Input
Range
With
Overranging
Typical
Gain Error at
15 to 35
°
C
(% of Reading)
Maximum
Gain Error at
–40 to 70
°
C
(% of Reading)
0–20 mA
4–20 mA
±
20 mA
0–24 mA
3.5–24 mA
±
24 mA
0.083%
0.083%
0.09%
0.32%
0.32%
0.35%