Chapter 2
Configuration and Installation
© National Instruments Corporation
2-11
GPIB-110 User Manual
Installation
After configuring both GPIB-110s and connecting the coaxial or fiber optic cable, follow these
instructions:
Verify the Voltage Requirement
The GPIB-110 is shipped from the factory with either a 100 to 120 VAC or 220 to 240 VAC
internal power supply. Verify that the voltage specified on the power supply matches the voltage
supplied in your area.
Caution:
Operating the unit at any voltage other than the one specified could damage
the unit.
1. Plug the utility power cord into an acceptable electrical outlet (100 to 120 VAC or 220 to 240
VAC, depending upon which model of the GPIB-110 you have). Then plug the other end of
the power cord into the back panel of the GPIB.
2. Link the first GPIB to the GPIB-110 with an appropriate cable (Type X1 or better). The
cable coming from the GPIB should be plugged into the back panel of the GPIB-110.
3. Link the second GPIB to the remote GPIB-110 with the appropriate GPIB cable.
Mounting
The GPIB-110 enclosure is designed for table top or rack-mount operation. Single- and dual-
unit rack-mount kits are available from National Instruments.
Power On
The GPIB-110 extension system is fully operational when power is applied to both units and
they are connected. If the extenders are set for IFC Wait Mode, it may be necessary to turn on
the System Controller last after the extenders and all other devices are operating. This step is
necessary if the System Controller executes only one IFC shortly after power on.
The preferred operating mode is to keep both extenders and at least two-thirds of the devices on
both buses turned on when there is any GPIB activity.
Self-Test Mode
The GPIB-110s are equipped with a Self-Test mode that determines if the GPIB-110 receivers,
transmitters, and packet transmission and reception circuitry are operating correctly. This mode
is controlled by Switch 4.
This mode should be used only if you want to check the operation of the GPIB-110. It is not
necessary to use this mode to operate the GPIB-110s.
Note: The self-test switch settings are reversed from the markings that appear on the switch.
To set your GPIB-110 for self-test mode, you must set the switch in the OFF position.