Input Requirements
Refer to the Figure B-41 for the M Series counter/timer circuitry.
Source Period and Pulse Width
Figure B-46 and Table B-30 show the timing requirements for Counter
n
Source. The
requirements depend on the synchronization mode.
Figure B-46.
Counter n Source Timing Requirements
Table B-29.
Selected Gate to Count Enable Delays
Time
Synchronization
Mode
Gating
Mode
Min (ns)
Max (ns)
t
4
80 MHz Source
Edge
0.5
5.0
Level
-1.0
0.5
Other Internal
Source
Edge
1/2 Source Period -
1 ns
1/2 Source
3 ns
Level
1/2 Source Period -
2.5 ns
1/2 Source Period –
1 ns
External Source
Edge
7.5
22.0
Level
6.0
18.0
Table B-30.
Counter n Source Timing
Time
Description
Synchronization
Mode
Min (ns)
*
Max (ns)
t
5
Counter
n
Source Period
80 MHz Source
12.5
—
Other Internal
Source
25.0
—
External Source
50.0
—
Counter n Source
t
5
t
6
t
6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com