4-28
|
ni.com
Chapter 4
Analog Input
Figure 4-19.
AI Convert Clock Too Fast For AI Sample Clock; AI Convert Clock Pulses Are
Gated Off
Figure 4-20.
AI Sample Clock and AI Convert Clock Improperly Matched;
Leads to Aperiodic Sampling
Figure 4-21.
AI Sample Clock and AI Convert Clock Properly Matched
It is also possible to use a single external signal to drive both AI Sample Clock and AI Convert
Clock at the same time. In this mode, each tick of the external clock causes a conversion on the
ADC. Figure 4-22 shows this timing relationship.
Figure 4-22.
One External Signal Driving Both Clocks Simultaneously
AI Sample Clock
AI Convert Clock
Sample #1
Sample #2
Sample #3
1 2 3
0
1 2 3
0
1 2 3
0
Channel Measured
AI Sample Clock
AI Convert Clock
Sample #1
Sample #2
Sample #3
1
2
3
0
0
Channel Measured
1
2
3
0
AI Sample Clock
AI Convert Clock
Sample #1
Sample #2
Sample #3
Channel Measured
1 2 3
0
1 2 3
0
1 2 3
0
AI Sample Clock
AI Convert Clock
Sample #1 Sample #2 Sample #3
1 2 3
0
1 2 3
0
1 …
0
Channel Measured