Figure 7. Laser Alignment Fixture Components
3
2
1
1. Allen Screw
2. Laser Alignment Fixture
3. Calibration Laser
5.
Use the alignment knobs on the measurement antenna fixture to move the measurement
antenna horizontally and vertically until the laser cross hits the center of the reference
antenna.
When the laser cross hits the center of the reference antenna, the measurement antenna is
aligned.
6.
Attach the reference antenna laser to the reference antenna fixture and plug the reference
antenna laser USB cable into the USB port inside the chamber.
7.
Use the mmWave OTA Validation Positioner Control Panel to move the laser until it is
aligned to the center of the cross marker on the measurement antenna fixture.
8.
When the laser cross hits the center of the cross marker, add
90°
to the current
Elevation
number and click the
Set Absolute Home
button on the
Settings
tab in the mmWave
OTA Validation Positioner Control Panel.
The position is saved in the chamber and remains after restarting the system.
9.
Unplug the calibration laser USB cable from the USB port.
10. Remove the laser alignment fixture from the measurement antenna fixture.