3.
In the
Load Configuration Dialog
dialog box, click the file button and browse to
<NIDocDir>
\OTA Test\System Calibration Files
.
4.
Select
OTA-Test-Cal-Base.cfg
and click
Open
.
Note
The calibration configuration assumes that the name of the PXIe-3622
in Measurement & Automation Explorer (MAX) is
mmWave_VST
.
5.
Select
Use existing calibration configuration
.
The
NI RF System Calibration Assistant Configurator
window launches.
6.
In the
Resources
window, verify that all resources and resource settings are correct for
the path loss characterization as shown in the following figure. Ensure the resource name
for the power meter matches the name used in MAX.
Figure 9. Resources Window
By default,
Self Calibrate Analyzer
and
Self Calibrate Generator
are not selected. NI
recommends that you follow the calibration recommendations in the
document and to perform self-calibration when the best accuracy numbers
are desired.
Note
NI does not recommend changing the default location of the calibration
files in the
File Resources
section.
mmWave OTA Validation Test Reference Architecture
|
© National Instruments Corporation
|
27