Refer to
Battery Replacement and Disposal
on page 16 for more information about replacing
the battery.
Opening and Closing the Door
Caution
Before opening the MON-10496 for service, you must turn off the
external power to the device and use the external breaker box to completely remove
mains power.
Attention
Avant d'ouvrir le MON-10496 pour l'entretien, vous devez couper
l'alimentation externe de l'appareil et utiliser le boîtier de disjoncteur externe pour
supprimer complètement l'alimentation secteur.
Notice
Do not put a load on the device door while it is open.
Complete the following steps to open and close the door of the MON-10496.
1.
Loosen the four captive screws on the door of the device with a 1/4 in. socket or 6 mm
(1/4 in.) flat-blade screwdriver.
2.
Open the door of the device slowly.
3.
Clean the gasket and gasket mating surface with a dry or damp cloth prior to closing the
device door.
4.
Close the door slowly.
5.
Tighten the four captive screws on the door of the device to 1.3 N · m (11.5 lb · in.).
Mounting
The MON-10496 must be mounted upright to a rigid surface such as a wall, panel, strut
channel, or metal post. For installations that may experience vibration, NI recommends using a
thread locking compound or lock nuts.
Each flange has several bolt locations as shown in the following figure. The flanges are 4 mm
(0.157 in.) thick. The V-notch on the rear facilitates mounting to a round surface such as a pipe
or pole using a pipe clamp bracket.