NI 5421/5441 Calibration Procedure
10
ni.com
Figure 2.
Optional Calibration Programming Flow
Self-Calibrate
Verify
Document
Pre-Adjustment
Results
Meets
Calibration
Test Limits?
Meets
Published
Specifications?
Update Calibration
Dates and
Temperatures
Without Adjusting
Results
Within
Published Specs
(Adjustment
Optional)
Adjust
Anyway?
Adjust
(Cal Dates and
Temperatures
Updated)
Verify
Document
Post-Adjustment
Results
Meets
Calibration
Test Limits?
Calibration/
Verification
Complete
Review
Verification/
Adjustment
Procedure or
Return Device
Yes
No
Yes
No
Yes
No
Yes
No