NI 5421/5441 Calibration Procedure
Table 4.
Values for Verifying the Gain of the Main Analog Path
Iteration
Gain
Ideal Positive
Full-Scale
(Volts)
Ideal Negative
Full-Scale
(Volts)
Calibration Test
Limit (Volts)
Published Spec
(Volts)
1
12.000000
12.000000
–12.000000
±0.019700
±0.048500
2
10.000000
10.000000
–10.000000
±0.016500
±0.040500
3
7.000000
7.000000
–7.000000
±0.011700
±0.028500
4
5.000000
5.000000
–5.000000
±0.008500
±0.020500
5
3.500000
3.500000
–3.500000
±0.006100
±0.014500
6
2.500000
2.500000
–2.500000
±0.004500
±0.010500
7
2.000000
2.000000
–2.000000
±0.003700
±0.008500
8
1.650000
1.650000
–1.650000
±0.003140
±0.007100
9
1.250000
1.250000
–1.250000
±0.002500
±0.005500
10
0.850000
0.850000
–0.850000
±0.001860
±0.003900
11
0.600000
0.600000
–0.600000
±0.001460
±0.002900
12
0.415000
0.415000
–0.415000
±0.001164
±0.002160
13
0.300000
0.300000
–0.300000
±0.000980
±0.001700
14
0.205000
0.205000
–0.205000
±0.000828
±0.001320
15
0.150000
0.150000
–0.150000
±0.000740
±0.001100
16
0.105000
0.105000
–0.105000
±0.000668
±0.000920
17
0.075000
0.075000
–0.075000
±0.000620
±0.000800
18
0.055000
0.055000
–0.055000
±0.000588
±0.000720
19
0.037500
0.037500
–0.037500
±0.000560
±0.000650
20
0.026000
0.026000
–0.026000
±0.000542
±0.000604
21
0.018500
0.018500
– 0.018500
±0.000530
±0.000574
22
0.013000
0.013000
– 0.013000
±0.000521
±0.000552
23
0.009000
0.009000
– 0.009000
±0.000514
±0.000536
24
0.006500
0.006500
– 0.006500
±0.000510
±0.000526
Note
:
Error Positive Full-Scale Value
= (
Measured Positive Full-Scale Value
) – (
Ideal Positive Full-Scale Value
)
Error Negative Full-Scale Value
= (
Measured Negative Full-Scale Value
) – (
Ideal Negative Full-Scale Value
)