National Instruments Corporation
43
NI 5421/5441 Calibration Procedure
5.
Call
niFgen_SetAttributeViInt32
to set the analog path value
(niFgen property node:
Output Attributes»Analog Path
) using
the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ANALOG_PATH
•
value
: The
Analog Path
value for the current iteration from
Table 9
6.
Call
niFgen_SetAttributeViReal64
to set the post-amplifier
attenuation (niFgen property node:
Calibration»Post-Amplifier
Attenuation
) using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_POST_AMPLIFIER_ATTENUATION
•
value
: The
Post-Amplifier Attenuation
value for the current
iteration from Table 9
Table 9.
Attributes and Values for the Main Path Post-Amplifier Gain and Offset
Iteration
Analog Path
Post-Amp
Attenuation
Current Configuration
1
NIFGEN_VAL_FIXED_LOW_GAIN_
ANALOG_PATH
0
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
LOW_GAIN_0DB
2
NIFGEN_VAL_FIXED_LOW_GAIN_
ANALOG_PATH
12
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
LOW_GAIN_12DB
3
NIFGEN_VAL_FIXED_LOW_GAIN_
ANALOG_PATH
24
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
LOW_GAIN_24DB
4
NIFGEN_VAL_FIXED_LOW_GAIN_
ANALOG_PATH
36
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
LOW_GAIN_36DB
5
NIFGEN_VAL_FIXED_HIGH_GAIN_
ANALOG_PATH
0
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
HIGH_GAIN_0DB
6
NIFGEN_VAL_FIXED_HIGH_GAIN_
ANALOG_PATH
12
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
HIGH_GAIN_12DB
7
NIFGEN_VAL_FIXED_HIGH_GAIN_
ANALOG_PATH
24
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
HIGH_GAIN_24DB
8
NIFGEN_VAL_FIXED_HIGH_GAIN_
ANALOG_PATH
36
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
HIGH_GAIN_36DB