NI 5421/5441 Calibration Procedure
44
ni.com
7.
Take the following voltage measurements at the NI 5421/5441 CH 0
front panel connector into a high-impedance load:
a.
Call
niFgen_SetAttributeViInt32
to set the offset DAC
value (niFgen property node:
Calibration»Offset DAC Value
)
using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OFFSET_DAC_VALUE
•
value
:
50000
b.
Call
niFgen_Commit
(niFgen Commit VI) to commit the
attribute values to the device using the following parameter:
•
vi
: The session handle returned from
niFgen_InitExtCal
c.
Wait 500 ms for the output to settle.
d.
Use the DMM to measure the voltage output by the device.
This measurement is
measurement 0
, which is used in step 8.
e.
Call
niFgen_SetAttributeViInt32
to set the offset DAC
value (niFgen property node:
Calibration»Offset DAC Value
)
using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OFFSET_DAC_VALUE
•
value
:
15000
f.
Call
niFgen_Commit
(niFgen Commit VI) to commit the
attribute values to the device using the following parameter:
•
vi
: The session handle returned from
niFgen_InitExtCal
g.
Wait 500 ms for the output to settle.
h.
Use the DMM to measure the voltage output by the device.
This measurement is
measurement 1
, which is used in step 8.