NI 5421/5441 Calibration Procedure
46
ni.com
4.
Call
niFgen_SetAttributeViBoolean
to set the analog filter state
(niFgen property node:
Output Attributes»Analog Filter Enabled
)
using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ANALOG_FILTER_ENABLED
•
value
:
VI_FALSE
5.
Call
niFgen_SetAttributeViReal64
to set the pre-amplifier
attenuation (niFgen property node:
Calibration»Pre-Amplifier
Attenuation
) using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_PRE_AMPLIFIER_ATTENUATION
•
value
:
0
6.
Call
niFgen_SetAttributeViReal64
to set the post-amplifier
attenuation (niFgen property node:
Calibration»Post-Amplifier
Attenuation
) using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_POST_AMPLIFIER_ATTENUATION
•
value
:
0
7.
Call
niFgen_SetAttributeViReal64
to set the output impedance
(niFgen property node:
Basic Operation»Output Impedance
) using
the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OUTPUT_IMPEDANCE
•
value
:
50
8.
Call
niFgen_SetAttributeViBoolean
to enable the analog output
(niFgen property node:
Basic Operation»Output Enabled
) using
the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OUTPUT_ENABLED
•
value
:
VI_TRUE