NI 5421/5441 Calibration Procedure
15. Call
niFgen_SetAttributeViBoolean
to disable the analog
output (niFgen property node:
Basic Operation»Output Enabled
)
using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OUTPUT_ENABLED
•
value
:
VI_FALSE
16. Call
niFgen_Commit
(niFgen Commit VI) to commit the attribute
values to the device using the following parameter:
•
vi
: The session handle returned from
niFgen_InitExtCal
17. Wait 500 ms for the output to settle.
18. Call
niFgen_ReadCalADC
(niFgen Read CAL ADC VI) to measure
the analog output voltage with the onboard calibration ADC using the
following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
numberOfReadsToAverage
:
3
•
returnCalibratedValue
:
VI_FALSE
•
calADCValue
: A ViReal64 variable. The variable passed by
reference through this parameter receives the voltage measured
by the onboard ADC. This value is
cal
ADC
measurement 0
,
which is used in step 32.
19. Call
niFgen_SetAttributeViBoolean
to enable the analog output
(niFgen property node:
Basic Operation»Output Enabled
) using the
following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OUTPUT_ENABLED
•
value
:
VI_TRUE
20. Call
niFgen_Commit
(niFgen Commit VI) to commit the attribute
values to the device using the following parameter:
•
vi
: The session handle returned from
niFgen_InitExtCal
21. Wait 500 ms for the output to settle.
22. Use the DMM to measure the NI 5421/5441 voltage output directly
into the DMM into a high-impedance load. This measurement is
external measurement 0
, which is used in step 32.