NI 5421/5441 Calibration Procedure
8
ni.com
Complete Calibration
Performing a complete calibration is the recommended way to guarantee
that the NI 5421/5441 meets or exceeds its published specifications for a
two-year calibration interval. At the end of the complete calibration
procedure, you verify that the output error falls within the calibration test
limits. Figure 1 shows the programming flow for complete calibration.
Figure 1.
Complete Calibration Programming Flow
Adjust
(Cal Dates and
Temperatures
Updated)
Verify
Document
Post-Adjustment
Results
Meets
Calibration
Test Limits?
Review
Verification/
Adjustment
Procedure or
Return Device
Verify
Self-Calibrate
Calibration/
Verification
Complete
Yes
No
Document
Pre-Adjustment
Results