background image

©

 National Instruments Corporation

11

NI 5762R User Guide and Specifications

Figure 5.  

NI 5762 - Getting Started (Host) VI Front Panel

Creating a LabVIEW Project and Running a VI on an FPGA Target

This section explains how to set up your target and create an FPGA VI and host VI for data 
communication. For more detailed information about acquiring data on your NI 5762R, refer to the 
device-specific examples available in NI Example Finder. 

Creating a Project

1.

Launch LabVIEW, or if LabVIEW is already running, select 

File»New

.

2.

In the 

New

 dialog box, select

 Project»Empty Project

 and click 

OK

. The new project opens in the 

Project Explorer

 window. 

3.

Save the project as 

5762SampleAcq.lvproj

.

Creating an FPGA Target VI

1.

In the 

Project Explorer

 window, right-click 

My Computer

 and select 

New»Targets and 

Devices

2.

In the 

Add Targets and Devices on My Computer

 dialog box, select the 

Existing Target or 

Device

 option button and expand 

FPGA Target

. The target is displayed.

3.

Select your device and click 

OK

. The target and target properties are loaded into the 

Project 

Explorer

 window.

4.

In the 

Project Explorer

 window, expand 

FPGA Target (RIO

x

, PXI-79

xx

R)

.

5.

Right-click 

FPGA Target (RIO

x

, PXI-79

xx

R)

 and select 

New»FPGA Base Clock

.

6.

In the 

Resource

 pull-down menu, select 

IO Module Clock 0

.

7.

Enter 

125 MHz

 in the 

Compile for single frequency

 control and click 

OK

.

Summary of Contents for NI 5762

Page 1: ...NI 5762...

Page 2: ...bVIEW project with the NI 5762R Contents Electromagnetic Compatibility Guidelines 2 How to Use Your NI FlexRIO Documentation Set 3 Front Panel and Connector Pinouts 4 Block Diagram 6 NI 5762 Component...

Page 3: ...nges or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules Caution To ensure the specified EMC perfor...

Page 4: ...LabVIEW FPGA Module Help Embedded in LabVIEW Help Contains information about the basic functionality of LabVIEW FPGA Module NI FlexRIO Help Embedded in LabVIEW FPGA Module Help Contains FPGA module a...

Page 5: ...evice and the chassis NI is not liable for any damage resulting from such signal connections For the maximum input and output ratings for each signal refer to the Specifications section of this docume...

Page 6: ...ata channel 4 DIO Port 0 2 Bidirectional single ended digital I O data channel 5 GND Ground reference for signals 6 DIO Port 0 3 Bidirectional single ended digital I O data channel 7 DIO Port 1 0 Bidi...

Page 7: ...Trigger Input Sync Clock IoModSyncClk PLL Locked Initialization Done Reinitialize AI 0 Data Over Range AI 1 Data Over Range 16 16 Configuration Error Configuration Error Code SPI Device Select SPI Re...

Page 8: ...integration functionality of the user defined CLIP but it also allows the CLIP to communicate directly with circuitry external to the FPGA Adapter module socketed CLIP allows your IP to communicate d...

Page 9: ...ports of four signals each and are accessed using a U8 data type and Boolean write enable signal The four PFI signals are accessed individually using Boolean controls The NI 5762 Multiple Sample CLIP...

Page 10: ...VIEW FPGA application This section explains how to use an existing LabVIEW FPGA example project to generate and acquire samples with the NI 5762R This example requires at least one SMA cable for conne...

Page 11: ...right click the Open FPGA VI Reference PXI 7952R function and select Configure Open FPGA VI Reference c In the Configure Open FPGA VI Reference dialog box click the Browse button next to the Bitfile b...

Page 12: ...ct and click OK The new project opens in the Project Explorer window 3 Save the project as 5762SampleAcq lvproj Creating an FPGA Target VI 1 In the Project Explorer window right click My Computer and...

Page 13: ...information about configuring your clocks 15 In the Project Explorer window right click the FPGA target and select New VI A blank VI opens 16 Select Window Show Block Diagram to open the VI block dia...

Page 14: ...nd select Create Control to create a STOP button on the VI front panel window 10 Add the Read Write Control function located on the FPGA Interface palette inside the While Loop 11 Wire the FPGA VI Ref...

Page 15: ...ure 7 Figure 7 5762SampleAcq Host vi Block Diagram 19 Save the VI as 5762SampleAcq Host vi Running the Host VI 1 Connect one end of an SMA cable to AI 0 on the front panel of the NI 5762 and the other...

Page 16: ...an NI 5762 Typical values describe useful product performance that are not covered by warranty Typical values cover the expected performance of units over ambient temperature ranges of 23 5 C with an...

Page 17: ...imum voltage 3 V DC 5 Vpk pk AC Measurements Figure 8 Frequency Response Logarithmic Scale Table 5 Bandwidth 3 dB Device Lower 3 dB Frequency Upper 3 dB Frequency NI 5762 01 61 7 kHz 102 MHz NI 5762 0...

Page 18: ...ency Response NI 5762 01 only Logarithmic Scale Figure 10 Frequency Response NI 5762 02 only Logarithmic Scale 80 90 100110120 Frequency MHz dBFS 0 1 2 3 4 5 1 5 10 20 30 40 50 60 70 NI 5762 01 6 0 2...

Page 19: ...Figure 12 Single Tone Spectrum at 20 MHz 3 1 dBFS SFDR 90 8 dBc NI 5762 01 and NI 5762 02 0 10 20 30 40 50 60 70 80 90 0 50 150 Frequency MHz dBFS 200 250 100 NI 5762 01 NI 5762 02 0 110 130 0 10 40 F...

Page 20: ...NR at 10 1 MHz 1 dBFS 74 6 dBc 1 dBFS 75 8 dBc SINAD at 10 1 MHz 1 dBFS 73 3 dBc 1 dBFS 75 2 dBc SFDR at 10 1 MHz with harmonics 6 dBFS 88 2 dBc 3 dBFS 88 4 dBc SFDR at 10 1 MHz without harmonics 6 dB...

Page 21: ...ernal stability 1 ppm CLK IN Number of channels 1 single ended Connector type SMA Input impedance 50 Input coupling AC Input voltage range 0 63 Vpk pk to 2 0 Vpk pk Absolute maximum voltage 3 0 V DC 3...

Page 22: ...out the supported analog input ranges and buffer currents refer to pages 19 and 23 of the AD9467 data sheet available at www analog com TRIG General Characteristics Number of channels 1 single ended C...

Page 23: ...ce with IEC 60068 2 1 and IEC 60068 2 2 Relative humidity range 10 to 90 noncondensing tested in accordance with IEC 60068 2 56 Altitude 2 000 m at 25 C ambient temperature Pollution Degree 2 Storage...

Page 24: ...er safety certifications refer to the product label or the Online Product Certification section Electromagnetic Compatibility This product meets the requirements of the following EMC standards for ele...

Page 25: ...ment visit ni com environment weee Where to Go for Support The National Instruments Web site is your complete resource for technical support At ni com support you have access to everything from troubl...

Page 26: ...e following instructions to install PXI EMC filler panels National Instruments part number 778700 01 in your PXI chassis 1 Remove the captive screw covers 2 Install the PXI EMC filler panels by securi...

Page 27: ...on your media or the National Instruments Patent Notice at ni com patents Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compli...

Reviews: