Appendix A
Specifications
A-2
ni.com
FIFO buffer size......................................512 samples
Data transfers ..........................................DMA, interrupts,
programmed I/O
DMA modes ...........................................Scatter-gather
(Single transfer, demand transfer)
Number of DMA channels .....................1
1
Configuration memory size ....................512 words
Accuracy Information
Transfer Characteristics
Relative accuracy....................................±1.0 LSB typ, ±3.0 LSB max
DNL ........................................................±0.5 LSB typ, ±1.0 LSB max
No missing codes....................................16 bits, guaranteed
Offset error
Pregain error after calibration..........±2.0
µ
V max
Pregain error before calibration.......±28.8 mV max
Postgain error after calibration ........±305
µ
V max
Postgain error before calibration .....±40.2 mV max
1
The NI 6013/6014 has one DMA channel to be shared by all resources on the device.
Nominal
Range at
Full Scale
(V)
Absolute Accuracy
Relative Accuracy
% of Reading
Offset
Noise + Quantization
(µV)
Absolute
Accuracy at
Full Scale
(mV)
Temp
Drift
Resolution (µV)
24 Hours
1 Year
(µV)
Single Pt.
Averaged
(%/
°
C)
Single Point
Averaged
±10
0.0658
0.0700
1897.5
933.0
82.40
8.984
0.0010
1084.90
108.49
±5
0.0158
0.0200
959.8
466.5
41.20
2.003
0.0005
542.45
54.245
±0.5
0.0658
0.0700
115.8
56.2
5.035
0.471
0.0010
66.299
6.630
±0.05
0.0658
0.0700
31.4
31.40
3.067
0.069
0.0010
40.382
4.038
Note:
Accuracies are valid for measurements after calibration. Averaged numbers assume dithering and averaging of 100 single-channel
readings. Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature and ±10 °C of
external or factory calibration temperature.