Appendix B
Timing Diagrams
B-12
ni.com
Figure B-11.
Sample Clock Timebase Timing and the Analog Input Timing Engine
Table B-5.
Convert Clock Timebase Timing
Time
Description
Line
Min (ns)
Max (ns)
t
13
Delay to Selected Start
PFI
3.4
8.8
RTSI
3.3
8.5
STAR
2.7
5.7
t
14
Selected Start Setup Time
(to Sync Convert Clock Timebase)
—
1.5
—
t
15
Selected Start Hold Time
(to Sync Convert Clock Timebase)
—
0
—
t
16
Sync Convert Clock Timebase to Start
—
0.9
2.4
t
17
Start to POUT
PFI
1.1
3.1
RTSI
1.1
2.7
Start
Terminal
Selected Reference Trigger
Reference Trigger
Terminal
Terminal
Selected Sample Clock
Terminal
Terminal
Terminal
Selected Start
RTSI
Terminal
Terminal
Terminal
Selected Pause Trigger
SI
Counter
Block
SI2
Counter
Block
SI_TC
Sample Clock Timebase
Sync Sample Clock Timebase
Convert Clock Timebase
Sync Convert Clock Timebase
SI Start
Pause Trigger
p_AI_Convert
Start
1
_i
SI2_TC