Chapter 7
Counters
©
National Instruments Corporation
7-39
In DAQmx, the device uses 80 MHz source mode if the user performs the
following:
•
Performs a position measurement
•
Selects duplicate count prevention
Otherwise, the mode depends on the signal that drives Counter
n
Source.
Table 7-6 describes the conditions for each mode.
80 MHz Source Mode
In 80 MHz source mode, the device synchronizes signals on the rising edge
of the source, and counts on the following rising edge of the source, as
shown in Figure 7-34.
Figure 7-34.
80 MHz Source Mode
Table 7-6.
Synchronization Mode Conditions
Duplicate
Count
Prevention
Enabled
Type of
Measurement
Signal Driving
Counter
n
Source
Synchronizati
on Mode
Yes
Any
Any
80 MHz Source
No
Position
Measurement
Any
80 MHz Source
No
Any
80 MHz
Timebase
80 MHz Source
No
All Except
Position
Measurement
20 MHz
Timebase, 100
kHz Timebase,
or PXI_CLK10
Other Internal
Source
No
All Except
Position
Measurement
Any Other
Signal (such as
PFI or RTSI)
External
Source
Source
Synchronize
Count