Appendix A
Specifications
A-8
ni.com
Physical
Dimensions (not including connectors)
NI PXI-783
x
R..................................16 cm by 10 cm (6.3 in. by 3.9 in.)
NI PCI-783
x
R..................................17 cm by 11 cm (6.7 in. by 4.3 in.)
I/O connectors.........................................Three 68-pin female high-density
VHDCI type
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth
.....................................±
12 V, Measurement Category I
Channel-to-channel
.................................±
24 V, Measurement Category I
Caution
Do
not
use the NI 783
x
R for connection to signals in Measurement Categories II,
III, or IV.
Environmental
The NI 783
x
R is intended for indoor use only.
Operating Environment
NI 7830R, NI 7831R
40 MHz or 80 MHz timebase ..........0
°
C to 55
°
C, tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2
NI 7833R
40 MHz timebase.............................0
°
C to 55
°
C, tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2
80 MHz timebase.............................0
°
C to 55
°
C except the
following: 0
°
C to 45
°
C when
installed in an NI PXI-1000/B or
NI PXI-101X; tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2