30
|
ni.com
|
NI 9144 User Guide and Specifications
Use the calibration coefficients with the following equation to generate corrected data:
NI 9205/9206
The vendor configuration extensions for the NI 9205/9206 are listed in Table 14.
Table 13.
NI 9203 Calibration Equation Information
Term
Units
Definition
I
corrected
pA
Calibrated current
I
raw
bits
The raw code from the NI 9203
I
expected0mA
bits
Expected code at 0 mA.
0 bits for 0-20 mA range.
32768 bits for ±20 mA range
LSB
weight
pA/bit
Number of pA in one bit
I
offset
pA
Offset at 0 mA
Table 14.
NI 9205/9206 Vendor Configuration Extensions
Index
Sub
Type
R/W
Description
0x2001
0
ARR:U32
—
Scan List = 33
1
R
Channels to Convert = <1..32>, default = 32
2..33
R/W
Channel Code
Icorrected Iraw
(
)
Iraw Iexpected 0mA
–
LSBweight
×
Ioffset
–
=
pA bits pA/bit pA