©
National Instruments Corp.
21
NI 9233 Operating Instructions and Specifications
Overvoltage protection (with respect to chassis gro
u
nd)
For an IEPE sensor connected
to AI+ and AI– ........................... ±30 V
For a low-impedance so
u
rce
connected to AI+ and AI– .......... –6 to 30 V
Inp
u
t delay
≤
25.65 kS/s ................................ 12.8/
f
s
+ 3
μ
s
>25.65 kS/s ................................ 9.8/
f
s
+ 3
μ
s
Acc
u
racy (–40 to 70 °C)
Calibrated typ ............................. ±0.1 dB
Calibrated max ........................... ±0.3 dB
Uncalibrated max ....................... ±0.6 dB
Acc
u
racy drift
Typical ........................................ 0.001 dB/°C
Maxim
u
m ................................... 0.0045 dB/°C