background image

NI 9265 Calibration Procedure

|

© National Instruments Corporation

|

3

Test Equipment

National Instruments recommends that you use the equipment in Table 1 for calibrating the 
NI 9265. If the recommended equipment is not available, select a substitute using the 
requirements listed in Table 1.

Test Conditions

The following setup and environmental conditions are required to ensure the NI 9265

 

meets 

published specifications.

Keep connections to the NI 9265 as short as possible. Long cables and wires act as 
antennae, picking up extra noise that can affect measurements.

Verify that all connections to the NI 9265 are secure.

Use shielded copper wire for all cable connections to the NI 9265. Use twisted-pair wire to 
eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ±5 °C. The NI 9265 temperature will be greater 
than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the NI 9265 measurement 
circuitry is at a stable operating temperature.

Initial Setup

Complete the following steps to set up the NI 9265.
1.

Install NI-DAQmx.

2.

Make sure that the NI cDAQ-9178 power source is not connected.

3.

Insert the module into slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of the 
NI cDAQ-9178 chassis empty.

4.

Connect the NI cDAQ-9178 chassis to your host computer.

5.

Connect the power source to the NI cDAQ-9178 chassis.

6.

Launch Measurement & Automation Explorer (MAX).

7.

Right-click the device name and select 

Self-Test 

to ensure that the device is working properly.

Table 1.  

Recommended Equipment

Equipment

Recommended 

Model

Requirements

DMM

NI 4070 DMM

Use a multiranging 6 1/2 digit DMM with a DC 
current measurement accuracy of 400 ppm.

Chassis

NI cDAQ-9178

Bench-Top

Power Supply

9 VDC to 36 VDC output voltage.

Summary of Contents for NI 9265

Page 1: ...s 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 5 Accuracy Adjustment 5 EEPROM Update 6 Reverification 6 Accuracy Under Calibration Conditions 6 Worldwide Support and Services 7 Software Calibrating the NI 9265 requires the installation of NI DAQmx 9 6 or later on the calibration system You can download NI DAQmx from ni com downloads NI DAQmx supports LabVIEW LabWindows CVI A...

Page 2: ...uctions and Specifications NI 9265 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming concepts and reference information about NI DAQmx VIs and functions NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties NI DAQmx NET Help Support ...

Page 3: ... 9265 temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9265 measurement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set up the NI 9265 1 Install NI DAQmx 2 Make sure that the NI cDAQ 9178 power source is not connected 3 Insert the module into sl...

Page 4: ...the DMM as shown in Figure 1 Figure 1 NI 9265 Current Accuracy Connections 3 Set the DMM to read DC current in the 20 mA range 4 Acquire a sample a Create and configure an AO current task on the NI 9265 according to Table 2 b Start the task c Generate a current output according to Table 3 d Wait the appropriate amount of time for the DMM measurement to settle e Read the NI 9265 output current meas...

Page 5: ...e DMM and the bench top power supply to the NI 9265 as shown in Figure 1 e Set the DMM to read DC current in the 20 mA range f Call and configure the NI 9265 setup calibration function with the DAC value obtained from the array of recommended calibration currents g Wait the appropriate amount of time for the DMM measurement to settle h Read the NI 9265 output current measurement from the DMM i Cal...

Page 6: ...ing your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the following conditions Ambient temperature 23 C 5 C NI 9265 installed in sl...

Page 7: ...d phone assistance from NI Application Engineers Visit ni com services for NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your National Instruments product Product registration facilitates technical support and ensures that you receive important information updates from NI A Declaration of Conformity DoC is our claim of compliance wi...

Reviews: