Continuous
60 V DC, Measurement Category I up to 5,000 m altitude
Withstand
500 V RMS, verified by a 5 s dielectric withstand test
Port-to-earth ground
Continuous
60 V DC, Measurement Category I up to 5,000 m altitude
Withstand
500 V RMS, verified by a 5 s dielectric withstand test
The maximum voltage that can be applied or output between any port or V
SUP
terminal and a COM terminal without creating a safety hazard.
Safety Compliance and Hazardous Locations Standards
This product is designed to meet the requirements of the following electrical
equipment safety standards for measurement, control, and laboratory use:
■
IEC 61010-1, EN 61010-1
■
UL 61010-1, CSA C22.2 No. 61010-1
■
EN 60079-0, EN 60079-7
■
IEC 60079-0, IEC 60079-7
■
UL 60079-0, UL 60079-7
■
CSA C22.2 No. 60079-0, CSA C22.2 No. 60079-7
Note
For safety certifications, refer to the product label or the
Certifications and Declarations
section.
Hazardous Locations
U.S. (UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, AEx nA IIC T4 Gc
Canada (C-UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, Ex nA IIC T4 Gc
Europe (DEMKO) Ex nA IIC T4 Gc