National Instruments Corporation
15
AO Waveform Calibration Procedure for NI-DAQmx
NI 6711/6713—12-Bit Resolution
NI 6722/6723—13-Bit Resolution
Table 4.
NI 6711/6713 Analog Output Values
Range (V)
Test Point
24-Hour Ranges
1-Year Ranges
Minimum
Maximum
Location
Value (V)
Lower
Limit (V)
Upper
Limit (V)
Lower
Limit (V)
Upper
Limit (V)
–10
10
0
0.0
–0.0059300
0.0059300
–0.0059300
0.0059300
–10
10
Pos FS
9.9900000
9.9822988
9.9977012
9.9818792
9.9981208
–10
10
Neg FS
–9.9900000
–9.9977012
–9.9822988
–9.9981208
–9.9818792
Table 5.
NI 6711/6713 Counter Values
Set Point (MHz)
Upper Limit (MHz)
Lower Limit (MHz)
5
5.0005
4.9995
Table 6.
NI 6722/6723 Analog Output Values
Range (V)
Test Point
24-Hour Ranges
1-Year Ranges
Minimum
Maximum
Location
Value (V)
Lower
Limit (V)
Upper
Limit (V)
Lower
Limit (V)
Upper
Limit (V)
–10
10
0
0.0
–0.0070095
0.0070095
–0.0070095
0.0070095
–10
10
Pos FS
9.9000000
9.8896747
9.9103253
9.8892582
9.9107418
–10
10
Neg FS
–9.9000000
–9.9103253
–9.8896747
–9.9107418
–9.8892582
Table 7.
NI 6722/6723 Counter Values
Set Point (MHz)
Upper Limit (MHz)
Lower Limit (MHz)
5
5.0005
4.9995