NI 5105 Calibration Procedure
|
© National Instruments
|
25
Note
The same time source is used for all channels, so you only need to verify the
timing accuracy on one channel.
Bandwidth
Table 8 contains the input parameters for verifying the bandwidth and the published
specifications of the NI 5105.
To verify bandwidth, complete the procedures described in this section for each of the
6 iterations listed in Table 8 for channel 0, then repeat for each of the remaining 7 channels on
the NI 5105.
Note
Verification of bandwidth is optional.
1.
Open a session and obtain a session handle using the niScope Initialize VI.
Table 7.
NI 5105 Timing Accuracy
Calibration Test Limits
Published Specifications
±5.3 ppm
±25 ppm
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the
following parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE