NI 5922 Calibration Procedure
|
© National Instruments
|
19
6.
Call
niScope_CalAdjustVCXO
(niScope Cal Adjust VCXO VI) with the following
parameter:
•
stimulusFreq
:
1,000,000
Note
The 1 MHz stimulus is automatically taken from channel 0.
7.
Call
niScope_CalAdjustRange
(niScope Cal Adjust Range VI) with the following
parameters:
•
channelName
:
"1"
•
range
:
0
•
stimulus
:
4.0
8.
Configure the DC source calibrator to output an exact -4 V DC signal into a 1 M
Ω
load
impedance (channel 1).
9.
Wait 3,000 ms for the calibrator to settle.
10. Call
niScope_CalAdjustRange
(niScope Cal Adjust Range VI) with the following
parameters:
•
channelName
:
"1"
•
range
:
0
•
stimulus
:
-4.0
11. Disconnect or disable all inputs to the oscilloscope.
12. Call
niScope_CalSelfCalibrate
(niScope Cal Self Calibrate VI) with the following
parameters:
•
channelList
:
VI_NULL
•
option
:
VI_NULL
13. Call
niScope_CalEnd
with the following parameters:
•
sessionHandle
: The instrument handle you obtained from
niScope_CalStart
•
action
:
NISCOPE_VAL_ACTION_STORE
to save the results of the calibration
You have finished adjusting the NI 5922. It is not necessary to adjust each channel individually.
However, you should repeat the entire verification procedure to verify a successful adjustment.