Chapter 2
Installation and Configuration
©
National Instruments Corporation
2-13
Request for All Status
Note
Voltage units are in V
DC
, and temperature units are in degrees Celsius.
Command
Response
Description
SA
<CR><LF>MONITOR,
209-2
xxx
-C
xx
-00 R
x
<CR><LF><CR><LF>
Header information
SP:PS:
x
<CR><LF>
Power supply status,
x
= OK or FAIL
ST:T1:
x
<CR><LF>
Temperature sensor #1,
x
= OK or FAIL
ST:T2:
x
<CR><LF>
Temperature sensor #2,
x
= OK or FAIL
ST:T3:
x
<CR><LF>
Temperature sensor #3,
x
= OK or FAIL
SV:V1:
x
<CR><LF>
+3.3VDC supply,
x
= OK or FAIL
SV:V2:
x
<CR><LF>
+5VDC supply,
x
= OK or FAIL
SV:V3:
x
<CR><LF>
+12VDC supply,
x
= OK or FAIL
SV:V4:
x
<CR><LF>
–12VCDC supply,
x
= OK or FAIL
SV:V5:
x
<CR><LF>
+5VDC VSB supply,
x
= OK or FAIL
SF:F1:
x
<CR><LF>
Fan #1,
x
= OK or FAIL
SF:F2:
x
<CR><LF>
Fan #2,
x
= OK or FAIL
SF:F3:
x
<CR><LF>
Fan #3,
x
= OK or FAIL
SF:F4:
x
<CR><LF>
Fan #4,
x
= OK or FAIL