background image

SAFETY, ENVIRONMENTAL, AND REGULATORY INFORMATION

NI PXI-2567

This document includes compliance precautions and connection information for the NI PXI-2567.

Note

 

The guidelines in this document are specific to the NI 2567. The other components in the system might not meet the same ratings. Refer to the

documentation for each component in the system to determine the safety and EMC guidelines for the entire system.

Safety Guidelines

Operate the NI 2567 only as described in the device user documentation.

Caution

 

Do not operate the NI 2567 in a manner not specified in the user documentation. Product misuse can result in a hazard. You can compromise

the safety protection built into the product if the product is damaged in any way. If the product is damaged, return it to NI for repair.

Safety Voltages

Connect only voltages that are within these limits.

Maximum drive voltage

50 VDC

Note

 

Overvoltage protection activates at 80 V maximum.

Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications.

These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational

electromagnetic environment.

This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a

peripheral device or test object, or if the product is used in residential or commercial areas. To minimize interference with radio and television reception and

prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation.

Furthermore, any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your

local regulatory rules.

Caution

 

To ensure the specified EMC performance, operate this product only with shielded cables and accessories.

Caution

 

To ensure the specified EMC performance, the length of all I/O cables must be no longer than 3 m (10 ft).

Preparing the Environment

Ensure that the environment in which you are using the NI 2567 meets the following specifications.

Environment

Operating temperature

0 °C to 55 °C

Storage temperature

-20 °C to 70 °C

Operating relative humidity

10% to 90%, noncondensing

Storage relative humidity

5% to 95%, noncondensing

Pollution Degree

2

Maximum altitude

2,000 m

Indoor use only.