SAFETY, ENVIRONMENTAL, AND REGULATORY INFORMATION
NI PXI-2567
This document includes compliance precautions and connection information for the NI PXI-2567.
Note
The guidelines in this document are specific to the NI 2567. The other components in the system might not meet the same ratings. Refer to the
documentation for each component in the system to determine the safety and EMC guidelines for the entire system.
Safety Guidelines
Operate the NI 2567 only as described in the device user documentation.
Caution
Do not operate the NI 2567 in a manner not specified in the user documentation. Product misuse can result in a hazard. You can compromise
the safety protection built into the product if the product is damaged in any way. If the product is damaged, return it to NI for repair.
Safety Voltages
Connect only voltages that are within these limits.
Maximum drive voltage
50 VDC
Note
Overvoltage protection activates at 80 V maximum.
Electromagnetic Compatibility Guidelines
This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications.
These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational
electromagnetic environment.
This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a
peripheral device or test object, or if the product is used in residential or commercial areas. To minimize interference with radio and television reception and
prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation.
Furthermore, any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your
local regulatory rules.
Caution
To ensure the specified EMC performance, operate this product only with shielded cables and accessories.
Caution
To ensure the specified EMC performance, the length of all I/O cables must be no longer than 3 m (10 ft).
Preparing the Environment
Ensure that the environment in which you are using the NI 2567 meets the following specifications.
Environment
Operating temperature
0 °C to 55 °C
Storage temperature
-20 °C to 70 °C
Operating relative humidity
10% to 90%, noncondensing
Storage relative humidity
5% to 95%, noncondensing
Pollution Degree
2
Maximum altitude
2,000 m
Indoor use only.