NI 5105 Calibration Procedure
Adjustment
If the NI 5105 successfully passed each of the verification procedures within the calibration test
limits, an adjustment is recommended but not required to guarantee the published specifications
for the next two years. If the device was not within the calibration test limits for each of the
verification procedures, you can perform the adjustment procedure to improve the accuracy of
the device. Refer to
to determine which procedures to perform.
An adjustment is required only once every two years. Completing the adjustment procedure
automatically updates the calibration date and temperature in the EEPROM of the device.
Complete all of the following steps to externally adjust the NI 5105.
1.
Obtain an instrument session handle using the niScope Cal Start VI.
Table 8.
NI 5105 Bandwidth Limits
Iteration
Max Input
Frequency
(MHz)
Range
(V)
Input
Frequency
(MHz)
Published Specifications
Max (dB)
Min (dB)
1
60
0.05
54.9
3
-3
2
60
0.20
59.9
3
-3
3
60
1.00
59.9
3
-3
4
60
6.00
59.9
3
-3
5
24
1.00
22.4
N/A
-3
*
6
24
1.00
29.9
-3
*
N/A
*
Published specifications value listed for validation of noise filter only. The specification is not included
in the device specifications document. Verification of these limits is optional.
LabVIEW VI
C/C++ Function Call
Call
niScope_CalStart
with the
following parameters:
resourceName
: The name of the device
to calibrate. You can find this name under
Devices and Interfaces in MAX
password:
The password required to
open an external calibration session. If
this password has not been changed since
manufacturing, the password is "
NI
"