NI PXIe-4463 Calibration Procedure
b.
Call
Basic Function Generator
using the parameters in Table 8.
c.
Write the resulting waveform to AO, and start the task.
4.
Measure the output voltage with the DMM.
5.
Stop the task.
6.
Repeat steps 2 to 5 for AO Attenuation of 17 dB and 37 dB. Refer to Table 9 for AO Gain,
DMM Range, and Signal Amplitude settings.
7.
Repeat steps 2 to 6 for pseudodifferential output terminal configuration.
8.
9.
Compare the DMM readings with the Max and Min limits in Table 10.
Table 8.
Basic Function Generator Parameters
Parameter
Value
Signal Type
Sine Wave
Frequency
1000 Hz
Amplitude
*
10 V
Sampling Info»Sampling Rate
51200 S/s
Sampling Info»Number of Samples
512
*
Refer to Table 9 for details about the correlation between AO Gain, DMM Range, and Signal
Amplitude.
Table 9.
Settings for AO Gain Verification
AO Attenuation (dB)
AO Gain (dB)
DMM Range (V)
Signal Amplitude (V)
0
0
10
10
17
-17
1
1.41421
37
-37
0.1
0.141421
Table 10.
AO Gain Verification Limits
AO Attenuation (dB)
As Found
Test Limits
As Left
Test Limits
Min (V
rms
)
Max (V
rms
)
Min (V
rms
)
Max (V
rms
)
0
7.0548
7.0874
7.06635
7.07578
17
0.9977
1.0023
0.99929
1.00070
37
0.09977
0.10023
0.099928
0.100071