NI PXIe-4463 Calibration Procedure
|
© National Instruments
|
13
AO Noise Verification
Complete the following procedure to verify AO noise.
1.
Connect channel AO 0 to all four NI PXIe-4464 inputs using the cables recommended in
Table 1. Use tee connectors to split the signal as shown in Figure 1.
Figure 1.
AO Noise Verification Setup
2.
Generate a voltage with the NI PXIe-4463:
a.
Configure an AO voltage channel task using the values in Table 15.
Table 15.
AO Setup
Configuration
Value
Physical Channels
Dev_name
/ao0
AO Gain
0 dB
Output Terminal Configuration
Differential
Rate
51200 S/s
Sample Mode
Continuous Sample
Samples Per Channel
512
NI PXIe-4464
DSA Analog Input
AI 0
AI 1
AI 2
AI 3
PFI 0
DSA Analog Output
NI PXIe-4463
PFI 0
AO 0
AO 1
ACCESS AO 0 AO 1
PFI 1