14
|
ni.com
|
NI PXIe-4463 Calibration Procedure
b.
Create an array of 512 points, and fill every element with value 0.
c.
Write the array to AO, and start the task.
3.
Measure the noise with the NI PXIe-4464:
a.
Configure an AI voltage channel task for the NI PXIe-4464 using the values in
Table 16.
b.
Start acquisition, and read data arrays for each input channel.
c.
Average the four input channels (add the arrays element by element, and divide by 4).
d.
Call
Standard Deviation and Variance
, and enter the previously computed
array. Obtain the noise as standard deviation output.
4.
Stop all previously started tasks.
5.
Repeat steps 2 to 4 for AO Attenuation of 17 dB and 37 dB.
6.
Repeat steps 2 to 5 for pseudodifferential output terminal configuration.
7.
8.
Compare the results with the limits in Table 17.
Table 16.
AI Setup
Configuration
Value
Physical Channels
Dev_name
/ai0:3
AI Gain
30 dB
Terminal Configuration
Differential
Excitation (IEPE)
0 A
Rate
49340 S/s
Sample Mode
Finite Number of Samples
Samples Per Channel
2467
Table 17.
AO Noise Verification Limits
Output Terminal
Configuration
AO Attenuation
(dB)
Noise Limits (Max)
(µV
rms
)
Differential
0
11.22
17
2.79
37
1.93
Pseudodifferential
0
11.42
17
3.18
37
2.60