NI PXIe-4463 Calibration Procedure
|
© National Instruments
|
3
Test Equipment
Table 1 lists the equipment required for the verification and adjustment procedures. If the
recommended model is not available, use equipment that meets the requirements listed in
Table 1.
NI-DAQmx C Reference Help
Reference information for NI-DAQmx C functions and NI-DAQmx C properties
NI-DAQmx .NET Help Support for Visual Studio
Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET
properties, key concepts, and a C enum to .NET enum mapping table
Table 1.
Required Equipment
Equipment
Recommended
Model
Where Used
Requirements
Digital
Multimeter
(DMM)
Keysight
Technologies
3458A
AO Offset
Verification
AO Offset and Gain
Adjustment
DC Voltage
Input Range: 1 V, 10 V
Accuracy:
2 ppm of range,
15 ppm of reading
AO Gain Verification
AO Flatness
Verification
AO Interchannel Gain
Mismatch Verification
AC Voltage
Input Range: 0.1 V
rms,
1 V
rms,
10 V
rms
Frequency Range:
20 Hz to 22.4 kHz
Accuracy:
45 ppm of range,
370 ppm of reading,
160 ppm of reading
@ 1 kHz