4
|
ni.com
|
NI PXIe-4463 Calibration Procedure
Digital
Multimeter
(DMM)
Keysight
Technologies
3458A
AO Current Drive
Verification
AC Current
Input Range: 0.1 A
rms
Accuracy:
200 ppm of range,
750 ppm of reading
Burden Voltage:
< 250 mV
Frequency
Counter
Keysight
Technologies
53220A-010
Timebase Frequency
Verification and
Adjustment
Resolution: > 10 digits
Timebase Accuracy:
±0.2 ppm
Analog Input
DSA
NI PXIe-4464,
BNC connectors
AO Noise Verification
Noise:
0.7 µV
rms
PXI Express
Chassis
NI PXIe-1082
All Tests
One of the following
chassis:
NI PXIe-1082
NI PXIe-1085
NI PXIe-1075
NI PXIe-1065
NI PXIe-1062Q
System
Controller
NI PXIe-8135
All Tests
A PXI Express controller
or MXI Express card
mXLR (F) to
BNC (M) Cable
(quantity 1)
NI 156789-0R46
or
NI 140150-0R46
All Tests
(mXLR Variant)
Length
0.5 m
BNC (M) to
BNC (M) Cable
(quantity 1)
Pomona
Electronics 5697
All Tests
(BNC Variant)
Length
0.5 m
BNC (M) to
BNC (M) Cable
(quantity 3)
Pomona
Electronics 5697
AO Noise Verification
Length
0.5 m
Table 1.
Required Equipment (Continued)
Equipment
Recommended
Model
Where Used
Requirements